BEAM
DATA FOR LINAC'S
When
taking Beam Data for LINAC's 2 chambers must be used:
- The
chamber inside the phanthom
- And
a chamber (monitor chamber) is placed somewhere outside the chamber
in the path of the beam
The
Chamber (monitor chamber) is placed outside the phanthom because
we need to stabalize the readings due to changes in the output from
the LINAC (the monitor chamber remains stationary and can be placed
anywhere in the beam)
The
2 readings are taken and then normalized to each other as shown
in the example below:

these
readings are normalized and then the new values are used to plot
the data.
WEDGES:
Hinge
Angle –
angle where 2 fields meet of a wedged pair according to the wedge
angle that is used.
Hinge(angle)
= 180 – 2 x wedge(angle)
Example
: using a 30 degree wedge
H(angle)
= 180 – (2 x 30)
H(angle)
= 180 – 60
H(angle)
= 120 (so there would be 120 degrees in between the wedged
fields)
CORRECTION
FOR COUNTOUR IRREGULARITIES:
3
Methods can be used to correct for irregulararities in a contour:
- Effective
SSD
- TAR
Method
- Isodose
Shift Method
Isodose
Shift Method:
We
will use shift factors to determine at what level the reading are
taken at:

this
data would be plotted as above and used to determine treatment fields
and dose inside the patient.
PARALLEL
OPPOSED FIELD DATA:
Readings
are taken at both sides of the phantom:
- There
will be a Dmax for the Anterior Beam and the Posterior Beam

The
data is plotted for the Anterior Beam and for the Posterior Beam
and then the readings are combined by adding the doses to the same
part of the phanthom on the


Now
the Isodose Curves are Drawn for the Combined Beams:

Now
we will Normalize the Plan to 100% to the Isocenter:
To
find the new values we will use the ratio:
Solving
for the new value:


And
for the Rest of the Values:
To
find the new values we will use the ratio:
Solving
for the new value:


As
seen above this Dose Matrix, the Isodose Curves have been normalized
to 100%, but in real life we like to use the following Isodose Curves:

Some
things to Note from the Graph above:
- Point
of normalization is the center of the phanthom and is 100%
- Maximum
dose in the phanthom (114%)
- Depth
of maximum dose (.5cm)
If
the tumor is at the center of the phanthom then:
In
the example below:
- The
tumor is receiving a minimum dose of 95% of the prescription dose
- The
treatment volume will extend from the Anterior Dmax point to the
Posterior Dmax point

The
Lateral Width of the Treatment Volume will be determined by the
field size and the Doctors definition of what level is included
in the treatment volume:
The
MD could prescribe to the 95% line or to the 90% line:

FOR
LINAC's THERE WILL BE A BETTER DOSE DISTRIBUTION:
- Better
can mean a lot of things but in this case we are saying the Dose
Distibution is more even with higher energy machines
- The
maximum dose in a patient with a Co60 machines is 114% but in
a 6MV LINAC is is 106%

|